The v0.3 schematic describes the active filter as a BandPass filter (72Hz - 720Hz).
If I calculate the poles of the filter I get 159.15 and 723.43, the lower value does not match, why is that? I found an online bode plotter http://sim.okawa-denshi.jp/en/optool.php and it agrees with the frequencies I calculate.
The text was updated successfully, but these errors were encountered:
Hey @kevinjwalters, we are using 2.2uF on the production board of Muscle BioAmp Shield. Thus you are seeing 72Hz-720Hz on the specifications. I will update the schematic to reflect this production change.
The v0.3 schematic describes the active filter as a
BandPass filter (72Hz - 720Hz)
.If I calculate the poles of the filter I get 159.15 and 723.43, the lower value does not match, why is that? I found an online bode plotter http://sim.okawa-denshi.jp/en/optool.php and it agrees with the frequencies I calculate.
The text was updated successfully, but these errors were encountered: