v1.5.3
- Plateau control system
- Telemetry
v1.5.2
- improve pressure control
v1.5.1
- calibrate pressure sensor's offset on startup
- better round displayed pressure values (in cmH2O)
- minor improvement to the pressure control
v1.5.0
- rework pressure and blower control
(blower will now take more time to ramp up/down but this will greatly improve the stability of injected air volume in many scenario) - rework alarms
(better pressure alarms, better battery alarms, better snooze behavior) - improve even more the measured and displayed pressures
(if no plateau pressure if found, screen will now display a?
instead of an uncertain value) - tweak the pressure control to make the plateau more accurate
- fix an issue with blower not restarting in some cases
- make sysclock more accurate
- improve code quality (MISRA)
- add a step in integration test to check O2 pipe
v1.3.2
- fix blower control
(it used to unexpectedly slow down a bit from times to times) - improve the measured and displayed pressures for peak and plateau
v1.3.1
- use the green LED near the start button to show whether the breathing mode is ON or not
- disable alarms related to the breathing cycle when program is stopped
- fix an issue with a battery alarm being briefly triggered at every boot
- integration test: open both valves at startup
v1.3.0
- add a program to test integration
- support electronic hardware v1 and v2
(v1 by default, v2 through a config flag) - implement start/stop
(from now on, machine will begin stopped!) - handle Emerson valves
(through a config flag) - improve the way buzzer is controlled
v1.2.3
- fix an issue with buzzer (sometimes it was stuck buzzing whereas no alarm were triggered)
- add battery-related alarms
v1.2.2
v1.2.2
v1.2.0
v1.2.0