Skip to content

manning-lab/metabolomics-qc

Folders and files

NameName
Last commit message
Last commit date

Latest commit

 

History

1 Commit
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Repository files navigation

Metabolomics QC for DoD project

Note that this pipeline is for reference rather than for use. Some code will not work out-of-the-box, as it includes code for wrangling data that was in a non-standard format.

The code that is the most reusable is 02_qc_mets.R, which accepts metabolite measurement data with metabolites as rows, samples as columns.

Data was not pushed to GitHub as it is restricted-access, but examples of what it looked like are given below.

Scripts

  • 00_install_pkgs.R: installs required R packages.
  • 01_preprocess_mets.R: Takes the raw data that was in non-standard format, extracts metabolite metadata, sample metadata, and metabolite measurements. Example below:
    • Raw data:
 |         |          |          |      |                           | Date extracted        | MM/DD/YY | MM/DD/YY
 |         |          |          |      |                           | Date injected         | MM/DD/YY | MM/DD/YY
 |         |          |          |      |                           | Column                | 1        | 1
 |         |          |          |      |                           | Analysis order        | 2        | 3
 | Method  | Compound | m/z      | RT   | HMDB ID (*representative) | Metabolite            | TOM###1  | TOM###2
 | HIL-pos | QI221    | 126.1369 | 7.45 | Internal Standard         | valine-d8             | 41368    | 42225
 | HIL-pos | QI197    | 174.1371 | 6.75 | Internal Standard         | phenylalanine-d8      | 136352   | 139558
 | HIL-pos | QI2      | 197.0677 | 3.58 | HMDB11103                 | 1.7-dimethyluric acid | 5141     | 648
  • Extracted metabolite metadata:
| Method  | Compound | m/z      | RT   | HMDB ID
| HIL-pos | QI221    | 126.1369 | 7.45 | Internal Standard         
| HIL-pos | QI197    | 174.1371 | 6.75 | Internal Standard         
| HIL-pos | QI2      | 197.0677 | 3.58 | HMDB11103                 
  • Extracted sample metadata:
extraction_date | injection_date | column | analysis_order | sample_id
MM/DD/YY        | MM/DD/YY       | 1      | 2              | TOM###1
MM/DD/YY        | MM/DD/YY       | 1      | 3              | TOM###2
  • Extracted measurement data:
           TOM###1 | TOM###2
QI221_hn | 41368   | 42225
QI197_hn | 136352  | 139558
QI2_hn   | 5141    | 649
  • 02_qc_mets.R: Performs QC on the formatted measurement data. Steps:
    • Remove metabolite with 0 variance
    • Remove metabolites with >25% missingness
    • Remove samples with >25% missingness
    • Apply various transformations to the data
    • Winsorize to the 5th standard deviation
    • Batch adjustment by median-scaling to a reference batch
  • 03_merge_mets.R: There were separate files for each LC-MS method (C8-positive, HILIC-positive, Amines, see here). This script merges the methods together.